GE DRAFT VOLUSON™ P8/VOLUSON™P6
DIRECTION 5775469, R
EVISION 3 DRAFT (JULY 19, 2018) BASIC SERVICE MANUAL
Chapter 10 - Care & Maintenance 10-17
Table 10-18 Type CF Patient Auxiliary Limits
Table 10-19 Isolation between Mains and Applied Limits****- Patient Leakage Main to Type BF
Applied part
Table 10-20 Isolation between Mains and Applied Limits****- Patient Leakage Main to Type CF
Applied part
Table 10-21 Total Patient leakage current: Type CF probe to ground (Only for Europe)
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 0.01 mA* 0.01 mA***
Close Protective earth and close neutral with reverse polarity 0.01 mA* 0.01 mA***
Open Protective earth and close neutral with normal polarity 0.05 mA* 0.05 mA***
Open Protective earth and close neutral with reverse polarity 0.05 mA* 0.05 mA***
Open neutral and close Protective earth with normal polarity 0.05 mA* 0.05 mA***
Open neutral and close Protective earth with reverse polarity 0.05 mA* 0.05 mA***
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 5 mA* 5 mA***
Close Protective earth and close neutral with reverse polarity 5 mA* 5 mA***
Conditions
Country
USA Others
Close Protective earth and close neutral with normal polarity 0.05 mA* 0.05 mA***
Close Protective earth and close neutral with reverse polarity 0.05 mA* 0.05 mA***
Conditions Limitation
Close Protective earth and close neutral with normal polarity 0.05 mA*****
Close Protective earth and close neutral with reverse polarity 0.05 mA*****
Open Protective earth and close neutral with normal polarity 0.1mA*****
Open Protective earth and close neutral with reverse polarity 0.1mA*****
Open neutral and close Protective earth with normal polarity 0.1mA*****
Open neutral and close Protective earth with reverse polarity 0.1mA*****