8-94 Section 8-41 - Replacing optional Peripherals
20.)Check the printer USB cable and power cable to passing through in EMI cover slot. Then
fasten 2 screws and assemble printer EMI cover 1 and 2.
21.)Route the B/W printer USB and power cable through the left side cable clamp.
Figure 8-114 Route the printer cables
22.)Fasten the 7 screws and fix the rear EMI cover.
23.)Assemble the rear cover, power cord and power cord bracket.
24.)Assemble the printer cover assembly.
Figure 8-115 Assemble printer cover assembly
EMI cover slot
Left side clamp
Fasten the 7 screws
Assemble the rear cover,
power cord and power
cord bracket
Printer cover assembly