GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-113
7.) Attach ECG label.
Figure 8-146 ECG label
8.) Perform: FRU8-47: Replacement of ECG Kit assy - Functional Tests
FRU8-2: System Software - Installation/Upgrade Procedure
FRU8-3: Software and Functional Checks after Installation/Upgrade Procedure
FRU8-4: Image Settings Only - Loading Procedure
go to 4-5-2 Load Image Settings Only
See Section
Functional and/or
Leakage Current Test
Debrief Script
4-3-1
Power On / Boot Up Proprietary Service Manual,
Voluson® P8 / Voluson®
P6
,
Revision Voluson® P8 /
Voluson® P6, Section 8-3.
Equipment passed all required
tests and is ready for use.
8-4
Software and Functional Checks after Installation/Upgrade Procedure
Table 8-1
System Function
4-3-2
Power Off / Shutdown
See Section
Functional and/or
Leakage Current Test
Debrief Script
4-3-1
Power On / Boot Up Proprietary Service Manual,
Voluson® P8 / Voluson®
P6
,
Revision Voluson® P8 /
Voluson® P6, Section 8-4.
Equipment passed all required
tests and is ready for use.
Table 8-1
System Function
4-3-2
Power Off / Shutdown
ECG label