4-36 Section 4-7 - Peripheral Checks
20.)Assemble the keyboard base using 4 flat washers, 4 spring washers and 4 hexagon
socket screws.
Figure 4-48 Assembling the keyboard base
21.)Re-assemble LCD cables through the LCD ARM, Refer to figure below.
Figure 4-49 Re-assembling LCD cables through the LCD ARM.
22.)Assemble HDMI cables for right direction, monitor power cable, monitor USB cable for left
direction to align the second colored point as shown. Then screw 1 screw to fix the cables
using MONITOR CABLE FIX BRACKET.
Figure 4-50 Routing the cables and screw 1 screw to re-assemble MONITOR CABLE FIX BRACKET