Installation Steps
19
Document No. PEXPERTIM - Rev. 1 - Release Date: May 2007
Configuration 2 - Mount to 16-inch center wood wall studs
Not included in shipping carton:
Mounting bolts appropriate for the wall support
Note! This configuration requires the optional Wall Plate (D800WPEXPERT).
1. Determine the proper location for mounting the unit. Referring to the
Mounting Template, verify that each electrical box is positioned
correctly in relation to location of the mounting studs.
2. Position the template or the Wall Plate itself at the appropriate height.
Mark the wall for the four mounting holes.
Note! If using the Wall Plate for marking, lay the plate aside before continuing.
Drill pilot holes appropriate for the mounting bolts that will be used.
Remove the Mounting Template (if used).
NOTE 2
NOTE 3
NOTE 1
NOTE 1
NOTE 1
NOTE 1
NOTE 5
NOTE 2
NOTE 4
NOTE 4
NOTE 3
MOUNTING WITH D800WP
SPECIAL 16" CENTER WALL PLATE
NOTE 1:
HOLE LOCATION TO MOUNT THE WALL PLATE VIA LAG BOLTS (SUPPLIED) TO WOODEN WALL STUDS SPACED AT 16" (40.6cm). MAXIMUM BOLT DIAMETER IS 10mm OR 3/8". BEFORE MOUNTING, VERIFY THAT WALL SUPPORT AND HARDWARE
MEET THE SUPPORT REQUIREMENTS LISTED ON THE OTHER SIDE.
NOTE 2:
THREADED HOLES TO MOUNT MASTER CONTROL BOX TO WALL PLATE WITH BOLTS SUPPLIED WITH WALL PLATE.
NOTE 3:
ELECTRICAL POWER ENTRY POINT. UNIT MUST BE CONNECTED TO ELECTRICIAN SUPPLIED POWER PER THE REQUIREMENTS LISTED ON THE OTHER SIDE. JUNCTION BOX POSITION SHOWN.
NOTE 4:
ENTRY POINT FOR THE INTERCONNECTING CABLE USED WHEN REMOTELY MOUNTING THE OPERATOR CONTROL PANEL. JUNCTION BOX POSITION SHOWN.
NOTE 5:
ALTERNATIVE ENTRY POINT FOR THE INTERCONNECTING CABLE USED WHEN REMOTELY MOUNTING THE OPERATOR CONTROL PANEL.
Route Control Cable Through Top Opening
Route Power Supply Wires Through Lower Opening
3. Place the Wall Plate back into position while routing the power supply
wires into the bottom opening. If the Master Control Touch Panel will
be mounted remotely, route the Control Cable (for the Remote Master
Control Touch Panel) into the top opening.
Figure 7 Route the Power Supply Wires and, if used, the
Control Cable
(Continued)