9-6
Gendex GXCB-500
TM
Service Manual
G990710 30 June 2008
RX Read X-ray Setup
See SX command for argument description.
0 = 120KV Range 1 = 90 KV Range
If no parameter is entered, then the command assumes 0.
SP Sets platform position - Sets the counts of the platform.
Must be less than 65000.
SR Sets rotation position - Sets the counts of the rotation.
Must be less than 733333.
SX Set X-ray
kvp, ma, frame rate, xrayon, xrayoff, panel readout,
panel_mode
Enter command as SX X1 X2 X3 X4 X5 X6 X7 X8 where
X is:
X1 specifies kV = 80 - 126 in increments of 2 only.
X2 specifies mA = 0-2. 0=3mA, 1=5mA, 2=7mA
X3 specifies frame rate = in between 50000000 - 833333
=> 50 MHz/frame rate (1-60 frames/sec)
X4 specifies X-ray on = typically set to 1 and between 0 -
(frame rate - 1)
X5 specifies X-ray off => greater than X-ray on and less than
(frame rate - 1) 5000 =1 msec (set as needed for desired
length)
X6 specifies Panel Readout = Typically set the same as the
X-ray off (except for continuous X-ray mode) in between
0 - (frame rate - 1)
X7 specifies Panel Pulse mode (0 or 1) (if not entered,
defaults to 0).
0 - 20 microsecond pulse for panel readout and valid frame.
1 - either a 5msec pulse for valid frame or 3msec for
panel readout
X8 specifies the tube 0 = 120 KV tube, 1 = 90 KV tube
Example: SX 120 1 3333333 1 780000 780000 0
If X7 and X8 are not specified, both are assumed to be zero.
X-Ray Commands
XP Frames
Maximum limit capped by the 44 second x-ray watchdog
Number of frames: 1 - 1024
Ethernet Watchdog 0 = disabled, 1 = enabled
(Number of frames*Frame Rate) <44 seconds