11-8
Gendex GXCB-500
TM
Service Manual
G990710 30 June 2008
Fallback Operations For VisionQ
If there is no network access for VisionQ, the best approach is to fix
the network. When the network is restored, click Retry on the No
Network Access screen and continue work as usual.
If network access cannot be quickly restored and scanning must
continue:
1. Click Cancel to fallback and use C:\LocalRoot.
2. Scan patients.
3. Exit VisionQ and restore the network as soon as possible. Any
changes made to studies such as workups, reports, or jpegs will
be swept to the server when the network connection is restored.
NOTE: Fallback is temporary until VisionQ exits unless the user
manually changes the setup, which is not recommended.
Fallback Operations For Vision
If there is no network access for standalone Vision, it is essential to
fix the network because C:\LocalRoot is most likely empty. When
the network is restored, click Retry on the No Network Access
screen and continue work as usual.
Network Fails During Operations
If a network failure occurs during use of VisionQ or standalone
Vision, the following occurs:
• The Study List goes blank (since it comes from the server
which is no longer accessible.)
• Error messages are displayed in the Sweeper log file.
• The Sweeper icon in the system tray of VisionQ changes to
the down state and a balloon is displayed for the state
change.
NOTE: Be aware that unsaved workups, reports, or jpegs may be lost
as a result of a network failure.