Safety Items
G990710 30 June 2008
Sample Site Plan
Below are two typical scanning room layouts that illustrate the
system interconnect for the Warning System, Operator Control Box
(Emergency Stop) and Interlock System which are all describe
above. Room layouts must provide a means for audio and visual
communication between the Operator and Patient. The Patient
Emergency Stop Box, which can stop the operation of the X-ray
device, must be within reach of the Patient when scanning occurs.
Scanner
Computer
Workstation
Operator Control Box
with 50 ft. (15.2m) Cable
Patient
Emergency Stop
Box with
10 ft. [3m] Cable
Interlock Cable
50 ft. [15.2m]
(if used)
Warning System
Cable
50 ft. [15.2m]
(if used)
CAT 5
Ethernet Cable
50 ft. (15.2m)
Hallway
Warning
Light or Alarm
(not supplied)
Room Size
6.5 x 8.5 feet [2 x 2.6 meters]
Interlock Switches
are not supplied
Window