2-8
Gendex GXCB-500
TM
Service Manual
G990710 30 June 2008
Interlock and Warning System Schematic
Scanner
Computer
Workstation
Operator Control Box
with 50 ft. (15.2m) Cable
Patient
Emergency Stop
Box with
10 ft. [3m] Cable
Interlock Cable
50 ft. [15.2m]
(if used)
Warning System
Cable
50 ft. [15.2m]
(if used)
CAT 5
Ethernet Cable
50 ft. (15.2m)
Hallway
Room Size
8.5 x 8.5 feet [2.6 x 2.6 meters]
Warning
Light or Alarm
(not supplied)
Interlock Switches
are not supplied
Customer
Connection
Back Panel
System Board
Interlock
System
Warning
System
Neutral
Return
Max. VAC 250V
Max. VDC 32 V
Max. Amps 1.5
Door Switch
Rating 12VD
100mA
Ground
Lamp
Ground Fuses
HIGH =
Door Close
Switch Close
OK to X-ray
HIGH
(X-ray ON)
Interlock
System
Warning System
Black
Red