6-4
Gendex GXCB-500
TM
Service Manual
G990710 30 June 2008
Patient Chair Alignment
To acquire quality images, the chair must be accurately leveled and
aligned.
Leveling Patient Chair:
1. From the Acquisition computer, double click the
Calibration icon.
The Calibration screen is displayed and the device
moves to the home position.
2. Insert Chair Center Locator into positioning block.
3. Press the Alignment Light button on the Patient Alignment
Panel to display lasers.
4. Check position of the
Horizontal and Vertical Lasers
relative to the notches on the
Chair Center Locator, as
shown.
If Horizontal Laser is out of
position, move Chair
Center Locator up or down
in the positioning block to
roughly align with
horizontal notch.
If the Vertical Laser is out of
position, then adjust the Chair
Foot, using a 1/4 inch nut driver.
Turn nut driver clockwise -
moves alignment tool towards
rear of Scanner.
Turn nut driver counter-
clockwise - moves alignment tool
towards front of Scanner.
Vertical
Laser
Horizontal
Laser
1/4
4