9-3
Troubleshooting and Adjustments
G990710 30 June 2008
IEC IC Command Codes
Home Commands
HB Home Beam Limiter
(HB = 0, 0, 0, 0 All four shutters completely open)
HFR Homes gantry and sets position to 100000
HP Homes gantry and sets position to 1500
HSR Quick home check if the machine is already homed
Movement Commands
MB Move Beam Limiter right left top bottom
Enter command as MB X X X X where X is:
(right left and bottom shutters range: 0 – 2600)
(top shutter range: 0 – 3300)
Example: MB 700 700 3000 1500
MP Move Platform [target] [speed]
Enter command as MP X X where X is:
(target range: 1500 – 100000, speed range: 1000 – 54000)
Example: MP 1500 1500
MPM Target Speed.
Enter command as MPM X X where X is:
Target or Position: between 0 - 90 mm (home position is zero
- absolute)
Speed: between 1000-54000
Example: MPM 20 1500
MPP Position type, moves receptor panel’s position.
Enter command as MPP X X where X is:
Position: 0 or 1 (Landscape or Portrait/Halfbeam)
Type: 0 or 1 (Platinum or Nano)
Example: MPP 1 0
MR Target Speed.
Enter command as MR X X where X is:
Target: between 35000 - 733333
Speed: between 1000 - 65000
Example: MR 35000 1000