EasyManua.ls Logo

Gendex GXDP-700 - Warnings and error messages

Gendex GXDP-700
136 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Using the unit
207723 rev 7
Gendex Dental Systems 69
5.6 Warnings and error messages
The unit responds to error situations by showing a dialog
box containing an error code and descriptive text on the
touch screen.
When an error code appears on the display the unit will
stop working and cannot be operated while the error code
is on the display. In less severe cases a warning message
will be displayed, leaving the unit operable.
5.6.1 Acknowledging errors
Most errors may be acknowledged by closing the dialog
box the error is reported in. Some errors require the unit to
be rebooted. If such an error occurs, or if the unit fails to
operate as described in the user's manual, switch the unit
off, wait a few seconds and switch the unit on again.
5.6.2 Image transfer errors
If an image is not transferred successfully to the PC, close
and then reopen the dental imaging software and/or restart
the PC. DO NOT restart the unit as this will erase any
image that is stored in the unit memory and this retrievable
image will be lost. If restarting the PC and/or restarting the
dental imaging software does not allow you to retrieve the
images, contact technical support without restarting the
unit.
Approved: Laihonen Tuuli 2016-12-02 16:41
Reviewed: Nieminen Timo Antero 2016-12-02 15:31
Approved
See PDM system to determine the status of this document. Printed out: 2017-03-22 10:54:38
D507729, 7
Copyright © 2016 by PaloDEx Group Oy. All rights reserved.