41
Operator’s Manual 032-0176-EN Rev. 4
Technical Data
ORTHORALIX8500TUBEHEAD110-0184G2
ORTHORALIX8500COLUMNANDIMAGINGASSEMBLY110-0199G1,110-0199G3
COLLIMATOR 112-1143G1
Power supply
120V AC or 230V AC nominal
Frequency 50-60 Hz
Max. line voltage regulation at
120V
4% voltage drop max. during exposure
Max. line current rating
8.5A at 230V, 13.5A at 120V for 80kV and 10mA
Nominal line current (idle)
stand-by by (at 230V)1.1A
approx. (at 120V)
2.0A approx.
High voltage waveform true DC, via square-wave output electronic power converter
(50kHz) and voltage multiplier/rectier
HVDC Power Supply A regulated output switch-mode power supply. Outputs are
independent of line voltage uctuation.
60 kV to 80 kV, in 2 kV steps max. tolerance from nominal value ±5%
4 mA to 10 mA, steps 4,6,8,10 max. tolerance from nominal value ±5%
Max. deviation of
exposure time
+ 2%
Exposure Time Pan: 12 sec.
Pedo: 11 sec.
Tube Type
See Product Data Sheet on Installation Manual
Tube Housing
See Product Data Sheet on Installation Manual
Duty Factor 1:20
Focal Spot Indicators
3 raised indicators on Tubehead Cover
Output Dose Rate
approx. average 1.2 mGy/s at 70 kV, 8 mA, 500 mm.
Leakage radiation from the
tubehead:
less than 0.25 mGy/ hour at 1 m in any direction
Leakage Technique Factors
80 kV, .78mA
Weight
245 lbs, 115kg
Environmental Conditions
Operational Temperature: +10/+40°C
Storage Temperature: -40/+70 °C
Minimum Air Pressure: 50,000 Pa
Lasers
Class
2, conforms to IEC 60825-
1:2001
Wavelength 650nm
Beam Divurgence for
Collimated Beam
>1mrd
Pulse Duration 45 seconds
Maximum Output <1mW
Mains Source Resistance
Voltage Range (VAC) Max line resistance (ohm)
120 0.4 max
230 1.0 max
15