Do you have a question about the GenLab SDO/225/GDIG and is the answer not in the manual?
Remove all packaging material from between shelves and inner walls, and peel off outer protective film before use.
Check voltage/load, connect 240V units with supplied lead, or wire 110V units via qualified electrician.
Position shelves, switch on mains, oven heats to set point. Adjust set point using UP/DOWN buttons.
Set switch to 120°C for temps up to 110°C, or 200°C for temps above 110°C.
Red LED illuminates and disables heat if chamber temp exceeds overheat set point.
Check supply lead, plug, switch operation, and indicators before each use.
Maintain clean/dry condition. Check connections, fuse, overheat system, and control temperature periodically.
Clean outer surfaces and chamber with damp cloth. Access fan/element for debris removal via bottom cover.
| Temperature Range | Ambient +5°C to 225°C |
|---|---|
| Temperature Stability | ±0.5°C |
| Voltage | 230V, 50/60Hz |
| Material | Stainless steel interior, painted steel exterior |
| Safety Features | Over-temperature protection |
Loading...