Scrollto01FANSPEEDDRYERwithJ. Press S.
Change AUT to MAN with K. Set fan speed to 75 %.
8. Check the reading of the external differential pressure gauge. Enter the reading of
gauge (D) in CS 1000 and click Set High.
The reading should be between 200 and 400 Pa.
Ifitistoohigh,theltermaybeclogged;ifitistoolow,theremaybeaholeinthelter.
9. Check that the value (E) rises to the reading of the external differential pressure gauge.
10. Log off CS 1000.
11. Reinstate the fan.
Set fan speed to 0 %. Change MAN to AUT with K. Press S.
12. Log out.
13. Reinstate the differential pressure gauge.