A pre-use check must be performed after all types of cleaning.
If the pre-use check fails:
• Carefully shake/tilt the cassette (5-7 times)
• Run the cassette in a ventilator system with a test lung for 10 minutes
• Dry the cassette in room air (if no drying cabinet is available) 12-24 hours, depending
on surrounding conditions.
• Re-do the pre-use check
Cleaning the expiratory cassette
Pre-use check