Operating Manual MSA 330 /MSA 340 2 Fusion process
11
If the problem still remains, the fitting and the barcode shall be
replaced.
If all the steps above are not solving the issue, it is
recommended to get assistance from a Regional service
Center.
2.5.2 Manual insertion
Enabling the manual insertion mode (see “Configuration” chapter for
more details), allows the operator to select by himself, the fusion voltage
and time, instead of gathering these information from the barcode.
ENTER FUSION DATA
Voltage= 40 V
Fusion Time=00
00s
Esc OK
In this case, the user has to set by ()() the fusion voltage value, push
START/OK ( I ) to confirm the content and move to the next field. Then set
by ()() the least significant digit (seconds), move the cursor to the
next digits by ()() and set the other values, press START/OK ( I ) to
confirm the setting.
To cancel the operation, and go back to the previous page, the user
may press STOP/ESC (O) at any time.
To provide a more complete trace of the jointing, in the same way, the
operator can add the data concerning the fitting in use: manufacturer,
fitting type and diameter.
ENTER FUSION DATA
Manuf= +GF+
Type= I D=
20mm
Esc OK
When the insertion is finished, press START/OK ( I ) to move forward to the
fusion process.
2.6 Process incomplete pre-alert
Before starting the fusion process, if temperature of the unit is too high a
message occurs and it will be necessary to let the machine to cool
down!
INSTRUMENT TEMP.
LIMIT
=
65°C
Ti=70°C
221 V
01.01.08 15:21
55Hz