2-6
USER HANDBOOK
2.3 Operating Modes
Conned Space Monitor (CSM) Operation
In this mode the instrument operates as a safety monitor
for use when entering confined spaces which may
contain hazardous gas mixtures (ammable, toxic and/or
asphyxiant). The instrument continually samples the ambient
atmosphere to which the operator is exposed and generates
alarms for low oxygen, high oxygen, high ammable gas,
high toxic gas, Short Term TWA (STEL) and Long Term
TWA (LTEL).
During normal operation in CSM mode, the instrument
emits a condence beep and illuminates a pair of red LED’s
briey every 8 seconds. This function is programmable in
the instrument setup software.
The condence signal function makes the user aware that
the instrument is sensing for gas without the need for the
operator to constantly view the display:
The condence beep and / or LED indication can be disabled.
Combustible Gas Indicator (CGI) Operation
In this mode the instrument operates as a gas indicator
drawing a sample via a probe from points where gas is
suspected to be present. Alarms are disabled in CGI mode
as the instrument is generally used for measuring gas levels
rather than monitoring for the presence of gas.