2.11 Before Use Checks
The following checks should be carried out before every
operation:
The instrument is clean and in good condition.
The sample line and any accessories are in
good condition and leak free. Make sure that the
instrument displays a flow fault when the
instrument inlet / sample line is blocked. If a
ow fault is not displayed, check tightness of all
ttings.
Any accessory lters used are clean and in good
condition.
The batteries have sufficient power left for the
intended use of the instrument.
The instrument is still within the calibration period
and is calibrated for the gas that you are intending
to sample.
All gas ranges are operational and zeroed
correctly.