Gossen Metrawatt GmbH 13
SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
6 Characteristic Values
Complete characteristic values can
be found in the operating instructions
(see section 3 on page 8).
Reference Ranges
Line voltage 230 V AC 0.2%
Line frequency 50 Hz 2Hz
Waveform Sine (deviation between
effective and rectified
value < 0.5%)
Ambient
temperature +23 C 2K
Relative
humidity 40 60%
Load resistors Linear
Nominal Ranges of Use
Nominal line voltage 100 V … 240 V AC
Nominal line
frequency 50 … 400 Hz
Line voltage
waveform Sinusoidal
Temperature 0 … +40 °C
Ambient Conditions
Storage
temperature -20 C +60 C
Relative
humidity Max. 75%, no condensa-
tion allowed
Elevation Max. 2000 m
Place of use Indoors, except within
specified ambient condi-
tions
In order to avoid deviation due to excessive
temperature fluctuation, e.g. after transport
in low outdoor temperatures and subse-
quent operation in a warm indoor environ-
ment, it’s advisable to wait until the test
instrument has acclimatized before starting
any measurements.
If the test instrument is colder than the ambi-
ent air, condensation may occur at high
humidity, i.e. condensate may accumulate
on components. This could result in the
occurrence of parasitic capacitances and
resistances which, in turn, affect the mea-
suring circuit and measuring accuracy.
Power Supply
Supply network TN, TT or IT
Line voltage 90 V 264 V AC
Line frequency 50 Hz 400 Hz
Power
consumption 200 mA test: approx. 32 VA
10 A test: approx. 105 VA
25 A test: approx. 280 VA
Mains to test socket
(e.g. during
function test) Continuous max.
3600 VA, power is con-
ducted through the instru-
ment only,
switching capacity
16 A, ohmic load,
the AT3-IIS32 (Z745X)
adapter (for example) can
be used for current
> 16 A AC
Electrical Safety
Protection
category I per EN 61140
Nominal voltage 230 V
Test voltage 2.3 kV AC 50 Hz or
3.3 kV DC
(mains circuit / test socket
to mains PE terminal,
USB, finger contact, test
probe(s) test socket)
Measuring
category Designed for 300 V CAT II
(but reduced to 250 V
CAT II through the use of
fuses for increased user
safety. The user-friendly
fuses are replaceable and
replacements are easily
obtainable).
Pollution degree 2