EasyManua.ls Logo

Gossen MetraWatt ECUTEST ST BASE - Relevant Standards

Gossen MetraWatt ECUTEST ST BASE
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
14 Gossen Metrawatt GmbH
Safety
shutdown
At DUT differential current of
>10mA,
shutdown time: < 500 ms,
can also be set to > 30 mA,
in the event of probe current
(electronic fuse) during:
Leakage current measure-
ment: > 30 mA~/< 500 ms*
Protective conductor resis-
tance measurement:
> 250 mA~/< 1 ms
in case of continuous current
I>16.5A
Fuse
links Mains fuses:
2 ea. FF 500V/16A
Special fuse:
M 250V/250mA
10 A RPE test current
(feature G01) only:
1 ea. FF 500V/16A
Electromagnetic Compatibility
Product standard EN 61326-1
EN 61326-2-2
USB Data Port
Type USB slave for PC connection /
remote control**
Type 2 ea. USB master,
for data entry devices * with HID boot
inter
face,
for USB flash drive for data backup,
for USB flash drive for saving
reports as HTML files, for printers *
* See operating instructions for compatible test in-
struments
** Remote control only with extension: “Remote
Control via PC (IZYTRONIQ)” (included as stan-
dard feature with SECUTEST ST PRO and avail-
able with SECUTEST DB+ (Z853R or feature
KB01).
Bluetooth
®
data interface 2.1 + EDR
(test instruments with feature M01 only)
Frequency range Max. 2.5 mW (class II)
Transmission
intensity 2400 … 2483.5 MHz
Mechanical Design
Display 4.3" multi-display
(9.7 × 5.5 cm), backlit,
480 x 272 pixels with 24 bit
color depth (true color)
Dimensions
W × H × D:
295 ×145 × 150 mm
Height with handle: 170 mm
Weight SECUTEST ST BASE(10)/
PRO: approx. 2.5 kg
SECULIFE ST BASE25:
Approx. 4.0 kg
(depending on the test instru-
ment version)
Protection Housing: IP 40,
Test socket: IP 20
per EN 60529
Database
Number of
data records 50,000
1 data record = 1 DUT or
location node or cus-
tomer or individual mea-
surement
6.1 Relevant Standards
The test instrument has been manufactured
and tested in accordance with the following
safety regulations:
*
Firmware version 3.2.0 and lower: 12 mA
EN 61010-1 Safety requirements for electrical equipment
for measurement, control and laboratory use
– Part 1: General requirements
EN 60529 Test instruments and test procedures
Degrees of protection provided by enclo-
sures (IP code)
EN 61326-1 Electrical equipment for measurement, con-
trol and laboratory use – EMC requirements
– Part 1: General requirements
EN 61326-2 Electrical equipment for measurement, con-
trol and laboratory use – EMC requirements
– Part 2-2: Particular requirements – Test
configurations, operational conditions and
performance criteria for sensitive test and
measurement equipment for EMC unpro-
tected applications
EN 61557-16
Electrical safety in low voltage distribution
systems up to 1000 V AC and 1500 V DC –
Equipment for testing, measuring or moni-
toring of protective measures –
Part 16: Devices for testing the effectiveness
of protective measures of electrical devices
and/or electrical medical devices

Table of Contents

Related product manuals