SECUTEST ST BASE(10) / PRO and SECULIFE ST BASE(25)
30 Gossen Metrawatt GmbH
13.6 IPE – Protective Conductor Current
1
Connection also via VL2E, AT3 adapter, AT16DI/AT32DI
2
Only suitable for connecting the DUT to the test socket.
Not suitable for measurements with AT16DI or AT32DI
adapter.
Before conducting any leakage current measurements,
make sure that the “Ref. voltage L-PE” and “Alt. Test
Freq.” measurement parameters have been set correctly
in SETUP (see section 12.4).
Connect the DUT (to test socket).
Switch DUT on
Switch DUT off
Meas. Parameter Meaning
Measurement type
Direct Direct measuring method
(via test socket, AT16DI/AT32DI)
Differential Differential current measurement
(via test socket)
Alternative Equivalent leakage current
method
(via test socket
1
)
AT3 adapter
Only test instruments with feature
I01: measurement with AT3
adapter: AT3-IIIE, AT3-IIS or AT3-II
S32
See section 13.19
Clamp
Only test instruments with feature
I01:See section 13.18
Single fault (SFC) – only with direct measure-
ment type
Normal status Single fault simulation not active
N interrupted Fault simulation – only phase and
protective conductor are con-
nected to the DUT
2
Polarity – only with measurement type
direct, differential and AT3 adapter
Normal Selection of polarity for mains
voltage to the test socket
Reversed
Select measuring function
1
IPE
Select parameters
2
Set parameters
3
Connect the DUT
4
Direct / Diff / Alt
Start test
5
Acknowledge line voltage warning
6
!
Direct & differential & AT3 adapter:
Save measured values to clipboard
7
Stop test
8
Save measurements under ID no.
9