7-5 Calibration
= remote platform
= main scale
Note:
l Please calibrate the main scale first.
l The calibration steps of main scale and remote platform are the same. Press key to
switch the calibration for the main scale or remote platform before the zero point calibration.
l You can only do the calibration for the remote platform after you have set the max. capacity
and division of the remote platform.
Here take JCAII-6K as an example.
1. Press key and hold while turning on the scale. Do not release till
is displayed in the Unit Weight window.
3. With no load on the weighing pan, press key to start the zero point calibration and wait
till is displayed in the Unit Weight window. Input the first calibration weight value
by the numeric key (6K model, 1/3 of full load is 2kg). And then put the corresponding weights
on.
3... Press key and wait till is displayed in the Unit Weight window. And then put
the corresponding weights on.
4. Press key and wait till is displayed in the Unit Weight window. And then put
the corresponding weights on.