SRT 10/II
USER MANUAL
page 1 of 37
Sed Rate Timer 10/lI (SRT10/II)
including Software (G2M110BO, Vers. 1.1)
ESR Analyzer
(optional thermal printer available)
USER'S MANUAL
MAN-003 Revision 09
Revision date: January, 2019

Questions and Answers

  • S
    Sheila MartinezAug 14, 2025
    Why is my Greiner Bio-One Timer data result not credible?
    • A
      Ashley MooreAug 14, 2025
      The data result might not be credible if the sample has a clot, in which case you should repeat sample collection. Another reason could be that the sample has foam, so you should re-mix it gently. Other causes could be related to the time from sample collection, instructions for sample mixing, or automatic temperature conversion.
  • J
    joshuastewartAug 17, 2025
    What to do if no info on Greiner Bio-One Sed Rate 10/II display?
    • D
      David GreenAug 17, 2025
      First, check if the power is switched on. If it is, verify that the power supply is working; replace it if necessary. If the problem persists, it may be an internal issue requiring service assistance.
  • A
    Anthony MarshallAug 21, 2025
    What to do if Greiner Bio-One Sed Rate 10/II Timer data result is not printed?
    • R
      Robert KellyAug 21, 2025
      Check the printer's power supply and cable connection. You may also need to check the printer configuration, referring to the manual for guidance. Additionally, you can perform a printer test by powering it on while pressing the key f/f. If all else fails, consider replacing the printer.
  • S
    sbeardAug 25, 2025
    What to do if sample level is high or low on Greiner Bio-One Sed Rate 10/II?
    • J
      jasmineandersonAug 25, 2025
      If the sample level is either too high or too low, repeat the sample collection. If the label was not placed correctly, replace it and repeat the analysis.
  • A
    Amber DeanAug 28, 2025
    What to do if Greiner Bio-One Sed Rate 10/II Timer system stopped?
    • R
      robinsonanthonyAug 28, 2025
      The system may have stopped due to a motor or mechanical defect, so call service assistance.
  • A
    Amy WilliamsAug 31, 2025
    What does temperature error mean on Greiner Bio-One Timer?
    • B
      bjonesAug 31, 2025
      A temperature error indicates a sensor malfunction. In this case, data analysis is not converted to 18°C, so call service assistance.
  • J
    Jeffrey SmithSep 3, 2025
    What to do if one or more question marks appear on Greiner Bio-One Sed Rate 10/II Timer display?
    • R
      Randy GuerraSep 4, 2025
      The appearance of question marks on the display may indicate an obstruction of the infrared barrier by external materials like label parts, or a disconnection/defect of the internal cable, so call service assistance.
  • M
    Matthew MossSep 6, 2025
    What to do if Greiner Bio-One Timer info on display scrambled?
    • V
      Valerie FergusonSep 6, 2025
      Call service assistance.
  • M
    Mike JohnsonSep 10, 2025
    What to do if sample has been removed from Greiner Bio-One Sed Rate 10/II Timer?
    • B
      Brian HarrisonSep 12, 2025
      Repeat analysis.

Summary

1. Introduction

1.1 Intended Use

Defines the primary purpose and application of the SRT 10/II ESR analyzer.

2. Potential Dangers and Safety Precautions

2.1 User Precautions

Essential safety rules for operators handling infectious materials and electro-mechanical systems.

2.2 Electrical Equipment Safety

Details potential electrical hazards associated with the instrument's power supply and components.

2.3 Mechanical Equipment Safety

Advises on precautions for the instrument's mechanical parts when powered on or off.

2.4 Biohazardous Material Handling

Guides safe handling and disposal of patient samples and QC products treated as biohazards.

3. Installation

3.3 Configuration Settings

Explains setting up the instrument via dip-switches for analysis time and results formats.

5. Operating Procedure

6. Temperature Compensation

6.1 Results Correction to 18°C

Explains automatic adjustment of results to a reference temperature of 18°C.

8. Error Information and Warnings

8.1 Warnings Information

Lists common warning messages displayed by the instrument and their meanings.

8.2 System Error Warnings

Describes critical system errors requiring service intervention.

9. Maintenance

9.1 Cleaning and Decontamination

Provides instructions for cleaning and decontaminating the instrument.

9.2 Periodic Test

10. Troubleshooting Guide

11. Technical Specifications

C. Decontamination Procedure

Greiner Bio-One Sed Rate Timer 10/II Specifications

General IconGeneral
Capacity10 samples
TypeSedimentation rate timer
ModelSed Rate Timer 10/II
ManufacturerGreiner Bio-One
DisplayDigital
UsageLaboratory