Section 3
3-30
3.7.2.2.2.1.3.<Probe Set> MENU Fn-3
(Mode_Edit_Probes_Setup_MAIN)
Probe set menu screen 1 of 2.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒Resistance RS: ▒
▒Serial No. RS: ▒
▒R0 Probe (or R(tpw)): ▒
▒Ser. No. Probe: ▒
▒RRate: ▒
▒Test Current (mA): ▒
▒──────────┬──────────┬─────────┬──────────▒
▒Edit │Ok │ │Screen 2 ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Press <Edit> and use arrow keys to move the highlight to the data line to be
changed. Modify data and press <Ok>.
A resistance value >0 for the ice point resistance (R0) or triple point of water
resistance (Rtpw), MUST be entered for the probe before a measurement can
take place.
<Probe Set> MENU Fn-3 (Mode_Edit_Probes_Setup_MAIN)
Probe set menu screen 2 of 2.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒Max Current (mA): ▒
▒ ▒
▒ ▒
▒ ▒
▒ ▒
▒ ▒
▒──────────┬──────────┬─────────┬──────────▒
▒Edit │Ok │ │Screen 2 ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
Press <Edit> and use arrow keys to move the highlight to the data line to be
changed. Modify data and press <Ok>.