Section 3
3-32
Function Key <Fn-1> ,<Filter>, allows selection of the filter that will be used
on the displayed readings, where:
No Filter = display most recent reading
Decimation = display (100-N) percent of the
previous reading. Where N is the
percentage value set in the
<Sofcal><Filter> menu for
Decimation.
N Average = display the average of the last N
readings. Where N is the number of
readings to be averaged and is that
value set in the <Sofcal><Filter>
menu.
Function Key <Fn-2> ,<Power>, allows selection of a multiple of the power
to be dissipated in the test resistor as X1 (unity) or as X2 (two times power
level).
3.7.2.3. <Save> MENU Fn-3 (Resistors_Setup_MAIN)
This menu allows saving to memory of the resistor parameter information. Up to 15
resistor files can be saved. Screen 1 of 2.
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒
▒Resistors ▒
▒00 SN of Resistor___ 05 _________________ ▒
▒01 06 ▒
▒02 07 ▒
▒03 08 ▒
▒04 09 ▒
▒──────────┬──────────┬─────────┬──────────▒
▒Select │Ok │ │Screen 2 ▒
▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒▒