What to do if there is air flow but no water spray in my Guilin Woodpecker Medical Instrument AP-A Dental equipment?
- TthomasjohnsonAug 2, 2025
If your Guilin Woodpecker Medical Instrument Dental equipment experiences air flow but no water spray, there are several potential causes: * The air source might be unconnected, or the air pressure might be too low (below 3 bar). Check the air connection and ensure the pressure meets the device requirements (5bar-7bar). * The handpiece could be blocked. Use a nozzle to dredge the handpiece. For other potential issues such as a blocked tail line, a blocked internal pipeline of the main unit, or a solenoid valve failure, it is recommended to contact local distributors or the manufacturer.