GDM-397/GDM-461 User Manual
49
GDM-461
Range
Resolution
Accuracy
Overload Protection
45~1kHz
>1kHz~10kHz
220μA
0.01μA
±(0.8%+10)
±(1.2%+50)
Fuse 1: F1A H 240V
(CE),φ6 x 25mm
2200μA
0.1μA
22mA
0.001mA
±(1.2%+10)
±(1.5%+50)
220mA
0.01mA
10A
0.001A
±(1.5%+10)
>1kHz~5kHz
Fuse 2: F10A H
240V (CE),φ6 x
25mm
±(2.0%+50)
Remarks:
When ≤ 5A: Continuous measurement is allowed.
When > 5A: Continuous measurement for less than 10 seconds with intervals of
more than 15 minutes between measurements.
True RMS is applicable from 10% of range to 100% of range.
AC crest factor can be up to 3.0 except 1000V (DC) where it is 1.5.
A residual reading of 10 digits with test leads shorted, will not affect stated
accuracy.