15
refrain from using the 20A terminal for more than 15
seconds when measuring high current.
Crest Factor Range
3.0 at full scale.
The burden voltage is the same as the DC current.
When the input exceeds the full scale of the selected range, the display will
indicate over-range: ―—OL―.
FREQUENCY MEASUREMENT AT ACA RANGE
RANGE
FREQUENCY
INPUT LEVEL (SINE
WAVE)
ACCURACY
500uA
10Hz ~ 20kHz
90μA
0.05%+1
5mA
10Hz ~ 20kHz
0.9mA
0.05%+1
50mA
10Hz ~ 20kHz
9mA
0.05%+1
500mA
10Hz ~ 20kHz
90mA
0.05%+1
2A
10Hz ~ 2kHz
1A
0.05%+1
20A
10Hz ~ 2kHz
9A
0.05%+1
AC+DC measurement does not support AC+Hz function.
Resistance
RANGE
RESOLUTION
ACCURACY
500
0.01
0.1%+4
5k
0.1
0.1%+2
50k
1
0.1%+2
500k
10
0.1%+2
5M
100
0.2%+2
20M
1k
0.3%+2
Open-circuit Voltage
3.2 volts maximum on 500Ω, 1.3 volts maximum on all
other ranges.
Protection
450V dc or peak ac continuous.
Capacitance
RANGE
RESOLUTION
ACCURACY
5nF *
0.001nF
1nF: 2%+10
<1nF & 0.5nF: 2%+20
50nF
0.01nF
10nF: 2%+10
<10nF & 5nF: 2%+30
500nF
0.1nF
2%+4
5uF
1nF
2%+4