Calibration Log
21
20 MΩ (19 MΩ
Range)
- 40 digits
______
+ 40 digits
Pass
Fail
DCV Calibration
Item
Min limit
Result
Max limit
Pass/Fail
400mV (500mV
Range)
- 3 digits
______mV
+ 3 digits
Pass
Fail
4V (5V Range)
- 3 digits
______ V
+ 3 digits
Pass
Fail
40V (50V Range)
- 3 digits
______ V
+ 3 digits
Pass
Fail
400V (500V
Range)
- 3 digits
______ V
+ 3 digits
Pass
Fail
1000V (1000V
Range)
- 3 digits
______ V
+ 3 digits
Pass
Fail
Diode Adjustment
Item
Pass
Fail
0.537V
Pass (0.6616 Vreading)
Fail
0.937V
Pass (1.0000 V reading)
Fail
DCA Adjustment
Item
Min limit
Result
Max limit
Pass/Fail
400uA (500uA
Range)
- 3 digits
______uA
+ 3 digits
Pass
Fail
4.0mA (5.0mA
Range)
- 3 digits
______ mA
+ 3 digits
Pass
Fail
40mA (50mA
Range)
- 3 digits
______ mA
+ 3 digits
Pass
Fail
400mA (500mA
Range)
- 3 digits
______ mA
+ 3 digits
Pass
Fail
2A (2A Range)
- 3 digits
______ A
+ 3 digits
Pass
Fail
8A (20A Range)
- 3 digits
______ A
+ 3 digits
Pass
Fail
ACA Adjustment
Item
Min limit
Result
Max limit
Pass/Fail
200uA/70Hz
(500uA Range)
- 4 digits
______uA
+ 4 digits
Pass
Fail