EasyManua.ls Logo

GW Instek GDM-8245 - Page 24

GW Instek GDM-8245
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GDM-8245 Service Manual
24
Short (100Ω
Range)
- 2 digits
______
+ 2 digits
Pass
Fail
Short (500Ω
Range)
- 3 digits
______
+ 3 digits
Pass
Fail
Short (5kΩ Range)
- 2 digits
______
+ 2 digits
Pass
Fail
Short (50kΩ
Range)
- 2 digits
______
+ 2 digits
Pass
Fail
Short (500kΩ
Range)
- 2 digits
______
+ 2 digits
Pass
Fail
Short (5MΩ
Range)
- 2 digits
______
+ 2 digits
Pass
Fail
Short (20 MΩ
Range)
- 2 digits
______
+ 2 digits
Pass
Fail
Resistance Range Verification
Item
Min limit
Result
Max limit
Pass/Fail
1Ω (1Ω Range)
- 4 digits
______
+ 4 digits
Pass
Fail
100Ω (100Ω
Range)
-11 digits
______
+ 11 digits
Pass
Fail
490 Ω (500Ω
Range)
- 37 digits
______
+ 37 digits
Pass
Fail
4.9 k(5kΩ
Range)
- 36 digits
______
+ 36 digits
Pass
Fail
49 kΩ (50kΩ
Range)
- 36 digits
______
+ 36 digits
Pass
Fail
490 kΩ (500kΩ
Range)
- 36 digits
______
+ 36 digits
Pass
Fail
4.9 MΩ (5MΩ
Range)
- 70 digits
______
+ 70 digits
Pass
Fail
19 MΩ (20 MΩ
Range)
- 41 digits
______
+ 41 digits
Pass
Fail
Diode Verification
Item
Pass
Fail
0.5~0.7
Pass
Fail

Table of Contents