GDM-8245 Service Manual
28
490mV/20kHz
(500mV Range)
______V
< 707 digits
Pass
Fail
490mV/50kHz
(500mV Range)
______V
< 1736 digits
Pass
Fail
AC+DC Verification
Item
Result
Max limit
Pass/Fail
200mV/50kHz
(500mV Range)
______V
< 721digits
Pass
Fail
ACV Verification (part II)
Item
Result
Max limit
Pass/Fail
4.9V/20Hz
(5V Range)
______V
< 353digits
Pass
Fail
4.9V/2kHz
(5V Range)
______V
< 182digits
Pass
Fail
4.9V/10kHz
(5V Range)
______V
< 353digits
Pass
Fail
4.9V/20kHz
(5V Range)
______V
< 707digits
Pass
Fail
4.9V/50kHz
(5V Range)
______V
< 1736digits
Pass
Fail
49V/20Hz
(50V Range)
______V
< 353digits
Pass
Fail
49V/2kHz
(50V Range)
______V
< 182digits
Pass
Fail
49V/10kHz
(50V Range)
______V
< 353digits
Pass
Fail
49V/20kHz
(50V Range)
______V
< 707digits
Pass
Fail
49V/50kHz
(50V Range)
______V
< 1736digits
Pass
Fail
490V/40Hz
(50V Range)
______V
< 353digits
Pass
Fail

Table of Contents