Calibration
63
Diode Calibration
Calibrator: 0.537V, 0.937V
GDM-8245 Range: Diode mode
1. Enter CL50 Calibration mode.
2. Set the mode to Diode. Press SHIFT + .
4. Check the DMM display. If the voltage is within the
reading range, go to the next calibration range.
Reading range: approx. 0.6616V
5. Press Auto/Man to re-calibrate the range. The mode
will change to CL51. Press Auto/Man again to save
when the calibration has settled. The mode will change
to CL52.
6. Repeat steps 3 to 5 with an input of 0.937V and a
reading range of 1V.
7. The mode will return to CL50.
Diode calibration is complete