5. Press Auto/Man to re-calibrate the range. The mode
will change to CL51. Press Auto/Man again to save
when the calibration has settled. The mode will return
to CL50.
6. Repeat steps 2 to 6 for the following inputs and ranges.
*Press SHIFT + ACA to enter the 20A range
The Up and Down arrow keys can be used to change the
DMM ACA range up and down.
Ensure the correct terminal is used for each range.
ACA calibration is complete
Capacitance Calibration
GDM-8245: 5nF, 50nF, 500nF, 5uF, 50uF
Multimeter Calibrator: 30nF, 300nF, 3uF, 30uF
Standard capacitance: 3.282nF
1. Enter CL50 Calibration mode.