GDM-8245 Service Manual
72
1. Set the range to 500uA (DCA mode).
2. Short the input terminals. Verify the results.
3. Repeat the procedure for the remainder of the ranges;
5.0mA~20A (DCA, DC 20A).
Note: Ensure the correct terminal is used for each
range.
DCA Verification (short) is complete
DCA Verification
(Digits): 490uA±70, 4.9mA±70, 49mA±70, 490mA±70,
1.9A±41, 19A*±41
GDM-8245: 500uA, 5.0mA, 50mA, 500mA, 2A, 20A
Calibrator: 490uA, 4.9mA, 49mA, 490mA,1.9A, 19A*
<20A*
*Do not input 19A for more than 1 second. Exceeding 1
second may blow the fuse.
1. Set the range to 500uA.
2. Input 490uA. Verify a DC current of 490uA±70 digits.