GDM-8245 Service Manual
74
3. Repeat for the remaining frequencies and ranges*.
Note: Ensure the correct terminal is used for each
range.
Specification verification is complete
ACV Verification (Short)
Accepted range
< 3 digits (all ranges)
Equipment
Test leads
Configurations
GDM-8245: 500mV, 5V, 50V, 500V, 1000V
Verification
procedure
1. Set the range to 500mV (AC mV).
2. Short the input terminals. Verify the results.
3. Repeat for the remaining ranges (AC mV and ACV).
ACV (shorting) verification is complete
ACV Verification (Part 1)
Accepted range
(Digits): 490mV(20Hz)<353; 490mV(2kHz)<182;
490mV(10kHz)<353; 490mV(20kHz)<707;
490mV(50kHz)<1736
Equipment
Multimeter calibrator
Test leads
Configurations
GDM-8245: 500mV
Calibrator: 490mV/20Hz/2kHz/10kHz/20kHz/50kHz
Verification
procedure
1. Set the range to 500mV (GDM-8245).
2. Input 490mV/20Hz. Verify the results.

Table of Contents