Verification
75
3. Repeat for the remaining frequencies.
ACV verification (part 1) is complete
AC+DC Verification
(Digits): 200mV(50kHz)<721 (AC + DC)
GDM-8245: 500mV range (AC+DC)
Calibrator: 200mV/50kHz
1. Set the range to 500mV (AC+DC)
2. Input 200mV/50kHz.Verify the results.
AC+ DC verification is complete
ACV Verification (Part 2)
(Digits): 4.9V(20Hz)<353; 4.9V(2kHz)<182;
4.9V(10kHz)<353; 4.9V(20kHz)<707;
4.9V(50kHz)<1736
49V(20Hz)<353; 49V(2kHz)<182; 49V(10kHz)<353;
49V(20kHz)<707; 49V(50kHz)<1736
490V(40Hz)<353; 490V(1kHz)<182
1000V(40Hz)<80; 1000V(1kHz)<45
800V(40Hz)<66; 800V(1kHz)<38
GDM-8245: 5V, 50V, 500V, 1000V
Calibrator: 4.9V/20Hz/2kHz/10kHz/20kHz/50kHz,
49V/20Hz/2kHz/10kHz/20kHz/50kHz,
490V/40Hz/1kHz, 1000V/40Hz/1kHz,
800V/40Hz/1kHz