GDM-8261 User Manual
166
VAL2?
Returns the 2
nd
display reading in the unit format specified in the
Configuration menu (Return Format, page 130) or from the
SYSTem:OUTPut:FORMat command (page 168).
Example: SAMP:COUN 100
VAL2?
>+0.345E-4,V DC
>+0.391E-4,V DC
> etc, for 100 counts.
Queries 100 counts of stored samples from the 2
nd
display.
TRIGger:SOURce
Selects the trigger source.
Parameter: INT | EXT
Example: TRIG:SOUR INT
Sets the trigger source as internal.
TRIGger:SOURce?
Returns current trigger source.
TRIGger:DELay
Sets the trigger delay in milliseconds
Parameter: <NRf>(0 ~ 9999) | MIN | MAX
Example: TRIG:DEL MAX
Sets the trigger delay to the maximum.
TRIGger:DELay?
Returns the trigger delay time in milliseconds.
Parameter: None | MIN | MAX
TRIGger:AUTO
Turns Trigger Auto mode on/off.
Parameters: ON | OFF
Example: TRIG:AUTO OFF
Turns the Trigger Auto mode off.
TRIGger:AUTO?
Returns the Trigger Auto mode.
Return parameter: 0|1, 0=OFF, 1=ON
SAMPle:COUNt
Sets the number of samples.
Parameter: <NR1>(1 ~ 9999) | MIN | MAX
Example: SAMP:COUN 10
Sets the number of samples to 10.