GDM-8261 User Manual
42
4. Select sensor
type
Press the Up/Down key. The
thermocouple type switches to the
next one.
5. Confirm and
go back to the
default display
Press the Enter key to confirm.
Press the Exit key to go back to the
default display.
Set Reference Junction Temperature (T-CUP)
Background
W
hen a thermocouple is connected to the GDM-8261,
the temperature difference between the thermocouple
lead and the GDM-8261 input terminal should be taken
into account and be cancelled; otherwise an erroneous
temperature might be added.
Type Range Resolution
SIM
(simulated)
0 ~ +50°C 0.01°C
The terminal temperature is manually defined by the user.
Default value: 23.00
1. Open
reference
junction menu
Press the Shift key, then the TEMP
(Sensor) key. The sensor selection
menu appears on the display.
TEMP
SENSOR
Press the Left and Right arrow keys
and select T-CUP (thermocouple).
Press Down, Right arrow key and
then Down again. The reference
junction selection menu appears on
the display.