GDM-8261 User Manual
64
Math Measurement
Applicable to
Background
Math measurement runs four types of mathematical
operations, MX+B, 1/X, Percentage and Stats, based on
the other measurement results.
Math type MX+B
Multiplies the reading (X) by the factor (M)
and adds/subtracts offset (B).
1/X
Inverse. Divides 1 by the reading (X).
Percentage
Runs the following equation.
(ReadingX – Reference)
x 100%
Reference
Stats
Performs standard deviation calculations
on measurement data.
Measure MX+B
1. Activate MX+B
Press the Shift key followed by the
2/4W (Math) key. The MX+B
setting appears.
2/4W
MATH
2. Set the factor
(M)
1st display
Shows the factor (M)
2nd display
Indicates MX+B (The letter M flashes)
1. Use the Left/Right key to move
the cursor (flashing point)
between the factor, digits, and
decimal point.