REMOTE CONTROL
173
CONFigure Commands
CONFigure:VOLTage:DC
Sets measurement to DC Voltage on the first display and specifies
range/resolution.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf>|
MIN | MAX | DEF)]
Example: CONF:VOLT:DC 1,MAX
Sets the voltage range to 1 volt and the resolution to the maximum.
CONFigure:VOLTage:AC
Sets measurement to AC Voltage on the first display and specifies
range/resolution.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf> |
MIN | MAX | DEF)]
Example: CONF:VOLT:AC
Sets the AC voltage range and resolution to auto range.
CONFigure:CURRent:DC
Sets measurement to DC Current on the first display and specifies
range/resolution.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf> |
MIN | MAX | DEF)]
Example: CONF:CURR:DC 10e-3,DEF
Sets the DC current range to 10mA using the default resolution.
CONFigure:CURRent:AC
Sets measurement to AC Current on the first display and specifies
range/resolution.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf>|
MIN | MAX | DEF)]
Example: CONF:CURR:AC 10e-2,MAX
Sets the measurement mode to ACI with a 100mA range at the maximum
resolution.
CONFigure:RESistance
Sets measurement to 2W Resistance on the first display and specifies
range/resolution.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf>|
MIN | MAX | DEF)]
Example: CONF:RES 10e3,MIN
Sets the range to 10kwith the lowest resolution.