REMOTE CONTROL
193
[SENSe:]DIODe:RESolution
Sets the Diode measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX)
Example:SENS:DIOD:RES 0.1e-4
Sets the Diode resolution to 0.00001
[SENSe:]DIODe:RESolution?
Returns the Diode measurement resolution.
Parameter: [None] | [MIN | MAX]
[SENSe:]TEMPerature:TCOuple:RESolution
Sets the thermocouple (T-CUP) measurement resolution. The resolution
depends on the rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX)
Example: SENS:TEMP:TCO:RES MAX
Sets the thermocouple resolution to the maximum.
[SENSe:]TEMPerature:TCOuple:RESolution?
Returns the thermocouple measurement resolution.
Parameter: [None] | [MIN | MAX]
[SENSe:]TEMPerature:FRTD:RESolution
Sets the 4W RTD measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX)
Example: SENS:TEMP:FRTD:RES MAX
Sets the 4W RTD resolution to the maximum.
[SENSe:]TEMPerature:FRTD:RESolution?
Returns the 4W RTD measurement resolution.
Parameter: [None] | [MIN | MAX]
[SENSe:]TEMPerature:RTD:RESolution
Sets the 2W RTD measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX)
Example: SENS:TEMP:RTD:RES MAX
Sets the 2W RTD resolution to the maximum.
[SENSe:]TEMPerature:RTD:RESolution?
Returns the 2W RTD measurement resolution.
Parameter: [None] | [MIN | MAX]