GDM-8261A User Manual
196
CALCulate:HOLD:REFerence?
Returns the percentage threshold from the Hold function.
CALCulate:REL:REFerence
Sets the reference value for the relative function.
Parameter: <NRf> | MIN | MAX
Example: CALC:REL:REF MAX
Sets the reference value to the maximum allowed.
CALCulate:REL:REFerence?
Returns the reference value from the relative function.
CALCulate:LIMit:LOWer
Sets the lower limit of the compare function.
Para meter: <NRf> | MIN | MAX
Example: CALC:LIM:LOW 1.0
Sets the lower limit to 1.0
CALCulate:LIMit:LOWer?
Returns the lower limit of the compare function.
CALCulate:LIMit:UPPer
Sets the upper limit of the compare function.
Para meter: <NRf> | MIN | MAX
Example: CALC:LIM:UPP 1.0
Sets the upper limit to 1.0
CALCulate:LIMit:UPPer?
Returns the upper limit of the compare function.
CALCulate:DB:REFerence
Sets the reference value for the dB function.
Parameter: <NRf> | MIN | MAX
Example: CALC:DB:REF MAX
Sets the reference voltage for dB measurements to the maximum allowed.
CALCulate:DB:REFerence?
Returns the reference voltage from the dB function.
CALCulate:DBM:REFerence
Sets the resistance value for the dBm function.
Parameter: <NRf> | MIN | MAX
Example: CALC:DBM:REF MAX
Sets the resistance value for dBm measurements to the maximum allowed.
CALCulate:DBM:REFerence?
Returns the resistance value from the dBm function.