REMOTE CONTROL
199
VAL1?
Returns the 1
st
display reading in the unit format specified in the
Configuration menu (Return Format, page 136) or from the
SYSTem:OUTPut:FORMat command (page 201).
Example: SAMP:COUN 100
VAL1?
>+0.333E-4,V DC
>+0.389E-4,V DC
> etc, for 100 counts.
Queries 100 counts of stored samples from the 1
st
display.
VAL2?
Returns the 2
nd
display reading in the unit format specified in the
Configuration menu (Return Format, page 136) or from the
SYSTem:OUTPut:FORMat command (page 201).
Example: SAMP:COUN 100
VAL2?
>+0.345E-4,V DC
>+0.391E-4,V DC
> etc, for 100 counts.
Queries 100 counts of stored samples from the 2
nd
display.
TRIGger:SOURce
Selects the trigger source.
Parameter: INT | EXT
Example: TRIG:SOUR INT
Sets the trigger source as internal.
TRIGger:SOURce?
Returns current trigger source.
TRIGger:DELay
Sets the trigger delay in milliseconds
Parameter: <NRf>(0 ~ 9999) | MIN | MAX
Example: TRIG:DEL MAX
Sets the trigger delay to the maximum.
TRIGger:DELay?
Returns the trigger delay time in milliseconds.
Parameter: None | MIN | MAX
TRIGger:AUTO
Turns Trigger Auto mode on/off.
Parameters: ON | OFF
Example: TRIG:AUTO OFF
Turns the Trigger Auto mode off.