REMOTE CONTROL
201
SYSTem Related Commands
SYSTem:BEEPer:STATe
Selects the beeper mode; no beep, beep on fail and beep on pass.
Parameter: <NR1>(0 | 1 | 2) 0=no beep, 2=fail, 1=pass
Example: SYST:BEEP:STAT 0
Turns the beeper off.
SYSTem:BEEPer:STATe?
Returns the beeper mode.
Return parameter: Beep on Pass | Beep on Fail | No Beep
SYSTem:BEEPer:ERRor
Sets the beeper to sound on an SCPI error.
Parameter: ON | OFF
Example: SYST:BEEP:ERR ON
Allows the beeper to sound when an SCPI error occurs.
SYSTem:BEEPer:ERRor?
Returns the beeper error mode.
Return parameter: 0|1, 0=OFF, 1=ON
SYSTem:ERRor?
Returns the current system error, if any.
SYSTem:VERSion?
Returns system version.
Return Parameter: X.XX.
SYSTem:DISPlay
Turns the Display on/off.
Parameter: ON | OFF
Example: SYST:DISP ON
Turns the display on.
SYSTem:DISPlay?
Returns the status of the display
Return parameter: 0|1, 0=OFF, 1=ON
SYSTem:OUTPut:FORMat
Sets the output format for the VAL1?, VAL2?, TRACe:DATA? and FETC?
queries. The measured value (V) can be set to be displayed with the
measurement units (U) and/or with the count number (C).
Parameter: <NR1>(0 ~ 3) 0=V, 1=V+U, 2=V+C, 3=V+U+C
Example: SYST:OUTP:FORM 3