ADVANCED MEASUREMENT
57
Refresh Rate
Function
S
M
F
(Readings/s)
Continuity / Diode
100
200
300
DCV/DCI/100~
100M (Accurate)
5
60
240
DCV/DCI/100~
100M (Quick)
30
600
2400
ACV/ACI
1.2
(sec/reading)
3.38
30
Frequency/Period
1
10
100
Selection steps
1. Press the Shift key followed by
the AUTO (RATE) key. The
refresh rate switches to the next.
/ EXITSHIFT
AUTO
RATE
2. The refresh rate indicator shows
the current status.
SMFS
Reading Indicator
Background
The reading indicator next to the 1st display flashes
according to the refresh rate when the captured data is
updated on the display.
Vm
Common Attribute: Manual/Automatic Triggering
Automatic
triggering
(default)
The GDM-8261A triggers according to the refresh rate.
See the previous page for refresh rate setting details.
Manual
triggering
Press the Trig key to trigger the
measurement manually. The trigger
must be set to external (EXT) for
manual triggering. See page 76.
TRIG