GDM-8351 User Manual
120
[SENSe:]TEMPerature:RJUNction:SIMulated
Set temperature simulation value.
Parameter: <NRf>(0.00 ~ 50.00)
Example: SENS:TEMP:RJUN:SIM 25.00
Sets the thermocouple junction temperature to 25˚C.
[SENSe:]TEMPerature:RJUNction:SIMulated?
Returns temperature simulation value.
Return parameter: <NR1> (+0000~+5000) ,where +0000=0.00˚C,
+5000=50.00˚C
[SENSe:]DETector:RATE
Sets the detection rate (sample rate)
Parameter: RATE(S | M | F)
Example: SENS:DET:RATE S
Sets the rate to slow (S).
[SENSe:]DETector:RATE?
Returns the sample rate.
Return parameter: SLOW, MID, FAST
[SENSe:]AVERage:TCONtrol
Selects the digital filter.
Parameter: MOV|REP
Example: SENS:AVER:TCON MOV
Sets the digital filter to the moving filter.
[SENSe:]AVERage:TCONtrol?
Returns the current digital filter type.
Return parameter: MOV(moving), REP(repeating)