EasyManua.ls Logo

GW Instek GDM-8351 - Additional Specifications

GW Instek GDM-8351
159 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GDM-8351 User Manual
150
Additional Specifications
The Additional Specifications apply in addition to the
Specifications listed on page 145 when the operating temperature
exceeds 18˚C ~ 28˚C.
DC Voltage
Measurement method: Sigma Delta A-to-D converter.
Input protection: 1000V peak on all ranges.
Range
Typical Input Impedance
100mV/1V
10.0 M±2% or >10G
10 V
11.1 M±2%
100 V
10.1 M±2%
1000 V
10.0 M±2%
Rate
Additional Rate Error Count
Med
50
Fast
200
DC Current
* 10mA~1A range has a 3V voltage limit protection and
F1.25A/1000V fuse protection.
And 10A range has a F12A/600V fuse protection.
Shunt resistance
Range
Shunt
Burden voltage
10mA
1
0.15V
100mA
1
1.5V
1A
0.1
0.8V
10A
0.01
0.6V
Rate
Additional Rate Error Count
Med
60
Fast
200

Table of Contents

Other manuals for GW Instek GDM-8351

Related product manuals