EasyManua.ls Logo

GW Instek GDM-8351 - Page 34

GW Instek GDM-8351
159 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GDM-8351 User Manual
34
For example:
A
Dynamic Range
0V
Dynamic Range
C D
B
E
A,B: Input exceeds the dynamic range.
C,D: The DCV offset causes the input to exceed the
upper dynamic range.
E: The DCV offset causes the input to exceed the
lower dynamic range.
The DC voltage range should be manually selected
when all of the following conditions are true:
When DCV measurement is used.
When the signals being measured contain both
DC and AC components.
When the amplitude of the AC component in
the measured signal is higher or lower than the
dynamic range of the range being currently
selected by the auto-range function.
Maximum DCV
Dynamic Range
Selected DCV Range
Dynamic Range
DC 100mV
±180mVmax
DC 1V
±1.8Vmax
DC 10V
±18Vmax
DC 100V
±180Vmax
DC 1000V
±1000Vmax

Table of Contents