OPERATION
39
Select the Current Range
The current range can be set automatically or manually.
Auto Range
To turn the automatic range selection On/Off,
press the AUTO key. The most appropriate
range for the currently used input jack will be
automatically selected. The DMM is able to do
this by remembering the last manually selected
range and using that information to determine
the smallest current range that the auto-range
function will switch to.
When the current input is switched to another
terminal, the range must be manually set.
Manual Range
Press the Up or the Down key to select the
range. The AUTO indicator turns Off
automatically. If the appropriate range is
unknown, select the highest range.
Selectable
Current Ranges
Range
Resolution
Full scale
INJACK
10mA
100nA
12mA
1A
100mA
1μA
120mA
1A
1A
100μA
1.2A
1A
10A
1mA
12A
10A
Note
For further details, please see the specifications on
page 145.